What are the common reasons for poor repeatability in measuring hardness values and large differences in multiple tests
Read:0
① The surface of the workpiece has oxide scale, oil stains, burrs, and has not been polished;
② Unstable placement of workpieces and suspended bottom;
③ Diamond indenter wear and chipping;
④ The equipment has not been leveled;
⑤ The indentation is too close to the edge of the workpiece, follow the indentation distance from the edge ≥ 2.5 times the diagonal of the indentation.